| Model Number |
5532-G-211 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.
|
| |
|
Event or Problem Description
|
|
Insert size is different.There should be an 11mm product when opened, but there was an estimated 9mm product.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding size/fit issue involving a triathlon insert was reported.The event was not confirmed.Method & results: product evaluation and results: the insert was received without its locking wire.Scratch marks and damage can be observed on the surface of an insert.Damage observed on the insert consistent with attempted implantation.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the insert size appears different.The event was not confirmed.The device was returned for evaluation.The insert was received without its locking wire.Scratch marks and damage can be observed on the surface of an insert.Damage observed on the insert consistent with attempted implantation.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
Insert size is different.There should be an 11mm product when opened, but there was an estimated 9mm product.
|
| |
|