Device has been returned and preliminary evaluation has been performed, however, our investigation is ongoing and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
As reported, during a endoscopic urethrotomy, upon opening the package of a universa soft ureteral stent set, the device was found to be broken.The break was found on the tethered end of the stent.The device was replaced to complete the procedure.No section of the device remained inside the patient's body.The patient did not require any additional procedures due to this occurrence.The complainant has not reported any adverse effects on the patient due to this occurrence.
|
|
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Summary of event: as reported, during a endoscopic urethrotomy, upon opening the package of a universa soft ureteral stent set, the device was found to be broken.The break was found on the tethered end of the stent.The device was replaced to complete the procedure.No section of the device remained inside the patient's body.The patient did not require any additional procedures due to this occurrence.The complainant has not reported any adverse effects on the patient due to this occurrence.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), and quality control procedures were conducted during the investigation.A visual inspection of the returned complaint device was also conducted.The complaint device was returned to cook in its original packaging, the device was broken at the 6th side port from the proximal tip.This type of damage is consistent with an unintended force on the tether.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows no other related complaints associated with the complaint device lot.Because there were no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in the field.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was functionally inspected by quality control and no notable gaps in production or processing controls were noted.There is no indication that a design or process related failure mode contributed to the reported event.Sufficient inspection activities are in place to identify this failure mode prior to distribution.The instructions for use (ifu) provides the following information to the user related to the reported failure mode: precautions improper handling can seriously weaken the stent.Acute bending or overstressing during placement may result in subsequent separation of the stent at the point of stress after a prolonged indwelling period.Angulation of the wire guide or stent should be avoided.Use of a 0-degree scope lens is recommended.Scopes larger than 21.0 french are suggested.Instructions for use endoscopic placement note: prior to use, immerse stent in sterile water or isotonic saline to allow the hydrophilic surface to absorb water and lubricious.This will ease placement under standard conditions.How supplied upon removal from package, inspect the product to ensure no damage has occurred.Based upon the available information and results of the investigation, cook has concluded that a definitive cause could not be determined.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|