• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CONTOUR ERCP CANNULA; ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) CANNULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CONTOUR ERCP CANNULA; ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) CANNULA Back to Search Results
Model Number M00530880
Medical Device Problem Codes Difficult to Advance (2920); Device Contamination with Chemical or Other Material (2944); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(initial reporter city, state): aizuwakamats, fukushima prefecture block h6: medical device problem code a180104 captures the reportable investigation results of foreign material present at the tip.Block h10: the returned contour ercp cannula was analyzed, and a visual evaluation noted that the device had white and black residue in the catheter.Additionally, there was a bend in the working length near the green label.A functional test was performed with two guidewires, navipro and novagold, and they passed through the distal tip with no problems.The unit was reviewed with the go pin gauge and it did not present any difficulty in the insertion.No other problems with the device were noted.The product analysis revealed there was an unknown white and black residue in the device.The residues could be from body fluids or a residue of the plastic material in the tip of the device; however, the exact nature of the residues is unknown.The reported problem of difficulty advancing the guidewire was not confirmed; however, during analysis it was found that the device was bent which could have contributed to the reported problem.It is likely that this could have been generated due to procedural factors such as handling technique or the force applied combined with resistance experienced during the procedure.Based on all available information, the conclusion code of the reported event is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported to boston scientific corporation that a contour ercp cannula was used in the common bile duct during a procedure performed on (b)(6), 2022.During the procedure, this contour ercp cannula was used with a non bsc guidewire during cannulation.When the contour ercp cannula was attempted to be pulled out leaving the guidewire in place, it was observed that there was difficulty removing the device; hence, the replacement of the device could not be performed.The procedure was completed another of the same device.It was reported that the guidewire was advanced without any problem.There were no patient complications reported as a result of this event.This event has been deemed a reportable event based on the investigation results: the product analysis revealed there was an unknown white and black residue in the device.Please refer to block h10 for full investigation details.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR ERCP CANNULA
Common Device Name
ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) CANNULA
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
LAKE REGION MEDICAL
31c butterfield trail
el paso TX 79906
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15152915
Report Number3005099803-2022-04231
Device Sequence Number5601297
Product Code ODD
UDI-Device Identifier08714729099833
UDI-Public08714729099833
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K833417
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/09/2023
Device Model NumberM00530880
Device Catalogue Number3088
Device Lot Number0006435347
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/08/2022
Initial Report FDA Received Date08/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-