| Model Number |
P6881CAP |
| Medical Device Problem Code |
Sparking (2595)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Hillrom received a report from a hillrom technician stating the power cord was damaged; it was sparking and it shorted out.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The hillrom technician found the power cable needed to be replaced.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the blower power cable to resolve the reported event.Based on this information, no further action is required.
|
| |
|