• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RFB, MDU, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RFB, MDU, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72200616F
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during set up of a knee procedure, the mdu had high temperature when connecting.There was no patient involvement.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).The reported device was received for evaluation.There was no relationship found between the device and the reported event.A visual inspection found a discolored cable.A functional evaluation revealed no problems.No overheating was noted.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Factors that could have contributed to the reported event include a failure of a concomitant device.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RFB, MDU, HAND CNTRL, PWRMX EL
Common Device Name
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15153249
Report Number1643264-2022-00272
Device Sequence Number12173961
Product Code HAB
UDI-Device Identifier03596010601209
UDI-Public03596010601209
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72200616F
Device Catalogue Number72200616F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2022
Supplement Date Received by Manufacturer08/31/2022
Initial Report FDA Received Date08/02/2022
Supplement Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-