BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ TRYPTICASE¿ SOY BROTH Back to Search Results
Model Number 221715
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that before using bd bbl¿ trypticase¿ soy broth was found to have bacterial contamination.The plates were not used.This is a report about contamination of the media.According to the customer's report, what appeared to be gram-positive bacillus was found in the media.The customer had experienced this issue before regardless of the lot.A gram staining was performed, however, it could not be stained.It is not growing.
 
Additional Manufacturer Narrative
H.6 investigation summary: material 221715 is manufactured by rehydrating the media components with usp purified water, and thoroughly mixing until a homogeneous solution is obtained.The tubes are filled, capped, torqued, and then labeled by machine per standard operating procedure (sop).The tubes are terminally autoclaved in an air over pressure (aop) autoclave, per manufacturing instructions, using a validated cycle.Post autoclaving, tubes are packaged into final shipping configurations.The batch history record review for batch 2026704 was satisfactory and no quality notifications were generated during manufacturing and inspection.Formulation, filling, torquing and autoclaving processes were within specifications.In process checks were performed during manufacturing at designated intervals per procedures.Those checks confirmed that the caps were tightened to the validated specifications per internal procedure.Qc inspection and testing were satisfactory at time of release.Also, each batch history record is reviewed to confirm the following prior to release: --the total elapsed time between end of formulation and start of the autoclave cycle was within the specified limits.--all autoclave parameters conformed to the validated cycle parameters for this product.--the minimum f0 for this product was met.The complaint history was reviewed and no other complaints for contamination have been taken on this batch.Retention samples from batch 2026704 (10 tubes) were available for inspection.No media defects were observed in 10/10 retention samples.All retentions tubes had the expected appearance for this product of light to medium light yellow, trace hazy to clear.For investigation, two retention tubes went into incubation.One retention tube was incubated in the 20-25 degree c incubator and one retention tube was placed in the 33-37 degree c incubator.At the end of the seventh day of incubation there were no signs of growth or turbidity.A gram stain was also performed, and no viable organisms were found.No photos were received to assist with the investigation.No returns were received to assist with the investigation.The complaint cannot be confirmed.Bd will continue to trend complaints for contamination.
 
Event or Problem Description
It was reported that before using bd bbl¿ trypticase¿ soy broth was found to have bacterial contamination.The plates were not used.This is a report about contamination of the media.According to the customer's report, what appeared to be gram-positive bacillus was found in the media.The customer had experienced this issue before regardless of the lot.A gram staining was performed, however, it could not be stained.It is not growing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL¿ TRYPTICASE¿ SOY BROTH
Common Device Name
BD BBL¿ TRYPTICASE¿ SOY BROTH
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15153687
Report Number1119779-2022-01052
Device Sequence Number7323447
Product Code JSG
UDI-Device Identifier30382902217153
UDI-Public30382902217153
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
PREAMENDMENT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/26/2023
Device Model Number221715
Device Catalogue Number221715
Device Lot Number2026704
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2022
Supplement Date Received by Manufacturer09/20/2022
Initial Report FDA Received Date08/02/2022
Supplement Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-