| Model Number |
CTR34 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Procedure performed: laparoscopic procedures event description: (b)(4).Date of event: (b)(6) 2022.The surgeon use the optical tecnic for the firt acces in laparoscopic procedures.During 2 different procedures the opturator of the trocar is too small to house the laparoscope.The surgeon inform that the 5 trocars had the same problem, but she don't remember the model and the lot of this trocars.Additional information received by email from applied medical representative on (b)(6) 2022: model ctr34 lot 1424056.Additional information received by email from applied medical representative on (b)(6) 2022: brand of laparoscope: [brand name] model of the laparoscope: 30° optics size: 10mm additional information received by email from applied medical representative on (b)(6) 2022: the user experienced inconsistent resistance during instrument manipulation.Patient status: no patient injury has been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The event device is anticipated to be returned to applied medical for evaluation.A follow-up report will be provided upon completion of the investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Testing was performed on the event unit, which confirmed the complainant¿s experience of improper fit.Upon closer inspection, the obturator was observed to have a misaligned feature inside the obturator.Based on the condition of the returned unit, the reported event was caused by the misaligned feature, caused by a molding defect.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.This event is not reportable as it is unlikely to cause or contribute to death or serious injury.Correction: updated section h6 component codes to 887 - mount.
|
| |
|
Event or Problem Description
|
|
Procedure performed: laparoscopic procedures.Event description: event 1 of 5: 2022-001526 (mfr report 2027111-2022-00679), event 2 of 5: 2022-001535 (mfr report 2027111-2022-00680), event 3 of 5: 2022-001536 (mfr report 2027111-2022-00681), event 4 of 5: 2022-001537 (mfr report 2027111-2022-00682), event 5 of 5: 2022-001538 (mfr report 2027111-2022-00683).Date of event: 30-31/05/2022.The surgeon use the optical tecnic for the firt acces in laparoscopic procedures.During 2 different procedures the opturator of the trocar is too small to house the laparoscope.The surgeon inform that the 5 trocars had the same problem, but she don't remember the model and the lot of this trocars.Additional information received by email from applied medical representative on 15june2022: model ctr34 lot 1424056.Additional information received by email from applied medical representative on 1july2022: brand of laparoscope: [brand name] model of the laparoscope: 30° optics, size: 10mm.Additional information received by email from applied medical representative on 12july22: the user experienced inconsistent resistance during instrument manipulation.Patient status: no patient injury has been reported.
|
| |
|
Search Alerts/Recalls
|