Model Number 100/189/085
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the blue 15mm connector kept disconnecting from the tracheal tube.The tip was not deep enough inside the tube and just disconnects.No clinical consequence, issue occurred upon opening.
 
Additional Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.Fifty-four (54) samples were returned from part number 100/189/075; sample was returned in unused conditions within original package.Visual inspection found the connector had properly assemble inside tube for 54 samples received.No other analysis was performed.Device passed all functional testing.No problem found.Root cause cannot be determined as the sample was successfully tested and no discrepancies were detected.This issue will be monitored for any increase in occurrence.If the occurrence of this issue increases significantly, additional investigative action will be taken accordingly.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLUE LINE SACETT SUCTION ABOVE THE CUFF TRA
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15154283
Report Number3012307300-2022-14402
Device Sequence Number5197072
Product Code BTR
UDI-Device Identifier15019315059551
UDI-Public15019315059551
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K081086
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 06/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/08/2024
Device Model Number100/189/085
Device Catalogue Number100/189/085
Device Lot Number3835859
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2022
Supplement Date Received by Manufacturer12/28/2022
Initial Report FDA Received Date08/02/2022
Supplement Report FDA Received Date06/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/08/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-