MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. TRACHEAL INTUBATION FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. TRACHEAL INTUBATION FIBERSCOPE Back to Search Results
Model Number LF-GP
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A user facility submitted a repair request to the olympus service center, on the tracheal intubation fiberscope for air/water leakage.Upon inspection and testing of the customer returned device, the coating of the scope insertion guide serpentine tube was found peeled off.(1-2mm or more).This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.There was no patient involvement, or user injury associated with this event.
 
Additional Manufacturer Narrative
The customer suspected device was returned for investigation.The olympus service center confirmed the reported event during inspection and testing.Upon evaluation of the returned device the following defects were found, water tightness not maintained due to breakage of the insertion tube, bending angle insufficient due to elongation of the angle wire, coat of the insertion tube came off, rupture of the flexible tube of the insertion site, liquid leakage observed in operation part, mark peeling of the operation part, curved rubber adhesive missing, scratches found on the eyepiece, scratches found on grip, scratches found on angle lever, scratches are found on the up/down plate, scratches found on the vent metal under the grip, mark peeling of the fixed ring, and peeling of the indicator on the base of the light guide mounting part.The faulty parts were replaced, and the device was returned to the user facility.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 17 years since the subject device was manufactured.Based on the results of the investigation, it¿s likely the reported event (coating of the insertion guide serpentine tube peeled) was due to stress of repeated use, external factors, or handling.The root cause of this event was unable to be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRACHEAL INTUBATION FIBERSCOPE
Common Device Name
TRACHEAL INTUBATION FIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-sh, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15154564
Report Number9610595-2022-00559
Device Sequence Number5208192
Product Code EOQ
UDI-Device Identifier04953170061998
UDI-Public04953170061998
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K981543
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLF-GP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/05/2022
Supplement Date Received by Manufacturer09/08/2022
Initial Report FDA Received Date08/03/2022
Supplement Report FDA Received Date10/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/24/2005
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-