Model Number SN6AT4
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
A healthcare professional reported following an intraocular lens implant procedure, scratch was observed on the iol inside the eye and decided to explant the iol and use the backup iol.Backup iol was implanted and the surgeon is complaining that the backup iol had on the surface some "rest" of an unknown material.He cleaned the surface, and it looks fine.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The lens was returned in a clear liquid inside a syringe.The lens was removed from the syringe for further evaluation.Both haptics are broken (possibly cut) from the optic.The optic was torn/split/cracked and cut into multiple pieces, typical of insertion and removal.The cut optic portions have multiple scratches, deep scrape marks, and scuff marks in the shape of an instrument used to grasp the lens on the anterior and posterior sides of the optic.The indicated video card was not received with the sample.A qualified cartridge and handpiece were indicated.The associated viscoelastic was not provided.It is unknown if a qualified product was used.The reported scratch was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company's release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The associated viscoelastic was not provided.It is unknown if a qualified product was used.The instruction for use (ifu) instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15154682
Report Number1119421-2022-01693
Device Sequence Number17451759
Product Code MJP
UDI-Device Identifier00380652250498
UDI-Public00380652250498
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSN6AT4
Device Catalogue NumberSN6AT4.235
Device Lot Number15356075
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/12/2022
Supplement Date Received by Manufacturer09/08/2022
Initial Report FDA Received Date08/03/2022
Supplement Report FDA Received Date10/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/21/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D.; MONARCH III IOL DELIVERY SYSTEM, INJECTOR.; UNKNOWN VISCOELASTIC.
Patient SexUnknown
-
-