Batch review performed on 28-jul-2022.Lot 2117638: (b)(4) items manufactured and released on 09-mar-2022.Expiration date: 2027-02-23.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported case during the period of the review.Additional devices involved: batch reviews performed on 03 august 2022: gmk-sphere 02.12.0512fr tibial insert fixed sphere flex size 5/12 mm r (k121416) lot 184232: (b)(4)items manufactured and released on 04-sept-2018.Expiration date: 2023-08-19.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event during the period of review.Gmk-sphere 02.07.1205r tibial tray fixed cemented size 5 r (k090988) lot 2118068: (b)(4) items manufactured and released on 25-mar-2022.Expiration date: 2027-03-10.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event during the period of review.
|
|
After the primary left knee surgery occurred on (b)(6) 2022, the post-op x-rays showed that right knee implants were implanted in the patient instead of left knee implants.The surgeon revised successfully the patient on (b)(6) 2022, with the correct left side implants.
|