SIEMENS HEALTHCARE GMBH-AT SENSIS VIBE HEMO; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 11007641
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code Cardiac Arrest (1762)
Date of Event 07/27/2022
Type of Reportable Event Death
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event or Problem Description
It was reported to siemens that an adverse event occurred while operating the sensis vibe hemo system.The user reported that the p3 pressure drifted up during the pci case.Additional information was provided that the patient passed away, however, no causal relationship was found between the failure and the death of the patient.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Additional Manufacturer Narrative
Siemens conducted a detailed investigation of the reported event.The investigation was performed on expert discussions (considering complaint description, cs (customer service) reports, and system history).In general, there are 4 pressure channels available (p1, p2, p3, and p4).These are used for invasive blood pressure monitor for diagnostic purposes.In case of the failure in any one of these pressure channels, the user can change the pressure transducer and continue the procedure within a minute.This issue can be and was identified by observing pressure waveform and can corrected by user immediately.This issue does not lead to worsening the patient condition and does not lead to death.In addition, the customer confirmed that there was no causal relationship between the patient's death and the identified sensis problem.Unfortunately, a more detailed investigation was not possible in this case because the customer cancelled an on-site inspection and repair.The occurrence rate of the reported error pattern was checked.A possible error accumulation or even a systematic error, which leads to a corrective action of the installed base, could not be determined by the investigation.The incident described in the adverse event is not classified as a reportable event after a thorough investigation because the death that occurred is not related to the product defect, nor is the product defect expected to cause serious injury, death, or unexpectedly long hospitalization to the patient, even if the defect occurs again.Internal id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSIS VIBE HEMO
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1- or-
rittigfeld 1
forchheim, germany 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1 or rittigfeld 1
rittigfeld 1
forchheim, 91301
GM   91301
Manufacturer Contact
rebecca tudor
40 liberty blvd.
65-1a
malvern, PA 19355
4843234198
MDR Report Key15156123
Report Number3004977335-2022-37948
Device Sequence Number5197118
Product Code DQK
UDI-Device Identifier04056869010199
UDI-Public04056869010199
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K150493
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number11007641
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2022
Supplement Date Received by Manufacturer09/20/2022
Initial Report FDA Received Date08/03/2022
Supplement Report FDA Received Date09/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexUnknown
-
-