• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S FOLYSIL CATHETER; CATHETER, RETENTION TYPE, BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S FOLYSIL CATHETER; CATHETER, RETENTION TYPE, BALLOON Back to Search Results
Model Number AA61081002
Medical Device Problem Codes Deflation Problem (1149); Difficult to Fold, Unfold or Collapse (1254); Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
According to the available information, in a neonatal intensive care unit, a nurse wanted to remove the urinary catheter from a new born.After deflation of the balloon, the nurse felt resistance during extraction of the last 0.5 centimeters; the catheter seemed impossible to remove.The catheter was introduced again, lubricated, then extracted with success, but with force.The catheter was tested once after removing it; the balloon correctly inflated but did not fold correctly during deflation, taking the aspect of a "ring".There was no clinical consequence for the newborn.
 
Additional Manufacturer Narrative
Lot number: 08136651.Should additional information prompt us to alter or supplement any information or conclusions contained in this report, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
A review of the complaint was performed and it was determined that the catheter was inflated with more liquid than was indicated by the valve.The lot # was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release and no trends were noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLYSIL CATHETER
Common Device Name
CATHETER, RETENTION TYPE, BALLOON
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebæk 3050
DA  3050
Manufacturer (Section G)
COLOPLAST A/S MANUFACTURING
9 avenue edmond rostand
sarlat-la-caneda 24206
FR   24206
Manufacturer Contact
usaby angela kilian
1601 west river road north
minneapolis, MN 55411
8007880293
MDR Report Key15156306
Report Number9610711-2022-00055
Device Sequence Number5183791
Product Code EZL
UDI-Device Identifier03600040126974
UDI-Public03600040126974
Combination Product (Y/N)N
PMA/510(K) Number
K013174
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAA61081002
Device Catalogue NumberAA6108
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/05/2022
Supplement Date Received by Manufacturer09/26/2022
Initial Report FDA Received Date08/03/2022
Supplement Report FDA Received Date10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-