| Model Number |
AA61081002 |
| Medical Device Problem Codes |
Deflation Problem (1149); Difficult to Fold, Unfold or Collapse (1254); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the available information, in a neonatal intensive care unit, a nurse wanted to remove the urinary catheter from a new born.After deflation of the balloon, the nurse felt resistance during extraction of the last 0.5 centimeters; the catheter seemed impossible to remove.The catheter was introduced again, lubricated, then extracted with success, but with force.The catheter was tested once after removing it; the balloon correctly inflated but did not fold correctly during deflation, taking the aspect of a "ring".There was no clinical consequence for the newborn.
|
| |
|
Additional Manufacturer Narrative
|
|
Lot number: 08136651.Should additional information prompt us to alter or supplement any information or conclusions contained in this report, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint was performed and it was determined that the catheter was inflated with more liquid than was indicated by the valve.The lot # was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release and no trends were noted.
|
| |
|