Model Number M0036181240
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during the pulmonary arteriovenous malformation embolization, the subject coil was used as the 11th coil.Physician felt resistance while advancing the subject coil from the introducer sheath and also at the tip of the microcatheter.He decided to retrieve the system, while doing so the subject coil detached prematurely at the proximal end of the microcatheter.The subject coil was taken out with the microcatheter.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.
 
Event or Problem Description
It was reported that during the pulmonary arteriovenous malformation embolization, the subject coil was used as the 11th coil.Physician felt resistance while advancing the subject coil from the introducer sheath and also at the tip of the microcatheter.He decided to retrieve the system, while doing so the subject coil detached prematurely at the proximal end of the microcatheter.The subject coil was taken out with the microcatheter.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.
 
Additional Manufacturer Narrative
H4 manufacturing date ¿ added.D4 expiration date ¿ added.Due to the automated manufacturing execution system (mes) there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, functional testing as well as visual testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information provided by the customer indicated that continuous flush set up and maintained throughout the procedure.No damage was noted to the packaging prior to opening, the device was confirmed to be in good condition during preparation/prior to use on the patient and was prepared as per the dfu.Friction was noted while the coil was advancing the introducer sheath.When an attempt was made to retrieve the catheter due to resistance at the tip of the microcatheter, the coil prematurely detached at the proximal end of the microcatheter.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable was assigned to this complaint of coil in catheter friction, main coil prematurely detached/separated during use and coil introducer sheath friction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XXL 360 12MM X 45CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15156880
Report Number3008881809-2022-00380
Device Sequence Number5268625
Product Code HCG
UDI-Device Identifier07613327128307
UDI-Public07613327128307
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K161429
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/18/2024
Device Model NumberM0036181240
Device Catalogue NumberM0036181240
Device Lot Number22869646
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/08/2022
Supplement Date Received by Manufacturer01/04/2023
Initial Report FDA Received Date08/03/2022
Supplement Report FDA Received Date01/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LEONIS MOVA HF MICROCATHETER (SUMITOMO BAKELITE)
Patient SexUnknown
-
-