Model Number V0100112
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Laceration(s) (1946); Unspecified Tissue Injury (4559)
Date of Event 07/18/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during a tka procedure, while using a visionaire adaptive guide kit - journey ii the anterior femoral cut was not flush, leaving the anterior flange off the cortex proximally.The tibia cut was very aggressive requiring an 18 mm poly to stabilize the joint spaces.Surgery was resumed, without any delay, with the same device.Patient outcome is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
Additional information added: h6 (component code, type of investigation, investigation findings, investigation conclusions).Results of investigation: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, the two undated, unlabeled x-ray photos show the gap between the femoral component and the femur that appears to contain cement.There is large joint space gap that is consistent with the reported 18mm insert.Based on a review of the documentation provided, the clinical root cause of the reported events could not be definitively concluded.It is unknown if and/or how this may have affected the joint line, leg-length, and if component itself could be impacted.It was reported that the procedure was resumed using the same device, without any delay.Since it was reported the patient¿s outcome is unknown, the impact to the patient beyond the unplanned over-resections, which required the use a thicker poly insert (upsized to 18mm), cannot be confirmed nor concluded based on the information provided.Therefore, no further clinical/medical assessment is warranted at this time.Should additional clinical documentation become available in the future, and/or if significant engineering evaluation findings are noted, the clinical/medical task may be re-evaluated.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for visionaire¿ patient matched instrumentation with fastpak instrument revealed that if the patient matched cutting block does not perform as intended, use the standard smith & nephew instrumentation to complete the surgery.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to surgical technique used, user/procedural variance or segmentation error.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Corrected data: h3, h6 (health effect - clinical code and health effect - impact code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIS ADPT GUIDE KIT JII
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15157053
Report Number1020279-2022-03596
Device Sequence Number5234563
Product Code JWH
UDI-Device Identifier00885556656419
UDI-Public00885556656419
Combination Product (Y/N)N
Initial Reporter StateAK
Initial Reporter CountryUS
PMA/510(K) Number
K170282
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberV0100112
Device Catalogue NumberV0100112
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2022
Supplement Date Received by Manufacturer09/02/2022
Initial Report FDA Received Date08/03/2022
Supplement Report FDA Received Date09/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-