| Model Number |
TCR10 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.During the visual analysis, the following was observed: the first photo shows a tyvek with the following information: tcr10 product code, 660a55 lot number and 2027-01-31 expiration date.The second photo shows a reload from the left view inside of a plastic bag with no apparent damage.Based on the two photos reviewed, the event described cannot be confirmed.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the instrument was not returned our evaluation is limited.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during intraoperative gastrectomy procedure, surgeon detected that the reload was not triggering all staples.There were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 10/17/2022.D4: batch # 379a57.Investigation summary: the product was returned for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample revealed that one tcr10 reload was received with no apparent damage.The reload was returned fully fired and with 3 b-formed staples on the deck.No functional test could be performed due to the condition of the reload.In addition, a tyvek was received along with the reload.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: place the instrument across the tissue for transection or into the lumen to form an anastomosis.For instruments with no knife, place the instrument across the tissue to be stapled.With the alignment/locking lever in the completely opened position, join the instrument halves together by aligning from either the front, center or back of the instrument.To adjust tissue on the forks before firing, move the alignment/locking lever to the intermediate position.This allows maneuvering of the tissue while the instrument halves are joined.Close the alignment/locking lever completely when the tissue is properly in place.To fire the linear cutter, place the thumb on the firing knob and two fingers on the shoulders of the linear cutter.Fire the instrument by pushing the firing knob completely forward.Completely return the firing knob to the original ¿return knob here¿ position.Separate the instrument halves by opening the alignment/locking lever and remove the instrument.Event described could not be confirmed as the reload was received fully fired.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|