• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B2120-020
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaint from this lot.The investigation was unable to determine a conclusive cause for the reported complaint.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of this device.
 
Event or Problem Description
It was reported after the 12x20mm armada 35 balloon dilatation catheter was removed from it's packaging the protective sheath could not be removed.Another device was used to successfully complete the procedure.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15157719
Report Number2024168-2022-08509
Device Sequence Number17442225
Product Code LIT
UDI-Device Identifier08717648155543
UDI-Public08717648155543
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model NumberB2120-020
Device Catalogue NumberB2120-020
Device Lot Number00118G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/19/2022
Initial Report FDA Received Date08/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-