ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B2070-040
Medical Device Problem Codes Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the procedure was to treat a lesion in the atrioventricular (av) fistula in the right arm, heavily calcified and heavily tortuous.The 7x40 mm armada 35 percutaneous transluminal angioplasty (pta) catheter was advanced to the target lesion and when inflated to 6 atmospheres (atm), the balloon ruptured.During removal of the catheter there was slight resistance with the sheath due to the rupture and the balloon separated in pieces.The physician had to widen the access site and with hemostats removed the balloon pieces successfully.The physician confirmed there were no pieces left in the patient's anatomy.A different access site was used and a new balloon catheter was used to complete the procedure.The vessel was patent with no embolism noted.The patient is doing well.No additional information was provided.
 
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
Visual analysis was performed on the returned device.The reported balloon rupture and separation were confirmed.The difficulty removing was not tested due to the condition of the device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation determined that the reported balloon rupture, difficulty removing, and separation resulting in unexpected medical intervention were likely due to circumstances of the procedure.It is likely that the balloon rupture was the result of interaction with the heavily calcified anatomy.There were no leaks reported during preparation, suggesting that the damage was not pre-existing.In addition, the difficulty reported during removal and separation of the balloon was likely the result of the ruptured balloon material catching on the introducer sheath during removal causing the balloon material to tear and separate radially.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15158088
Report Number2024168-2022-08516
Device Sequence Number20164217
Product Code LIT
UDI-Device Identifier08717648155352
UDI-Public08717648155352
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model NumberB2070-040
Device Catalogue NumberB2070-040
Device Lot Number91105G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/20/2022
Supplement Date Received by Manufacturer08/30/2022
Initial Report FDA Received Date08/03/2022
Supplement Report FDA Received Date09/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/05/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
Patient SexMale
Patient Weight74 KG
-
-