SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXGUARD FLOW CONTROLLER EXTENSION SET WITH NEEDLELESS Y-SITE(S); INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number MFS141
Medical Device Problem Codes Complete Blockage (1094); Leak/Splash (1354); Contamination of Device Ingredient or Reagent (2901)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd maxguard flow controller extension set with needleless y-site(s) experienced occlusion, leakage and foreign matter in fluid pathway.The following information was provided by the initial reporter: got an email from the doctor who said that the devices would not allow fluid to flow through despite being open.Water came out of the device.Unsure whether the device was being used for educational purposes.Not sure whether the water that came out was foreign matter or if it was part of the procedure.
 
Event or Problem Description
It was reported that the bd maxguard flow controller extension set with needleless y-site(s) experienced occlusion, leakage and foreign matter in fluid pathway.The following information was provided by the initial reporter: got an email from the doctor who said that the devices would not allow fluid to flow through despite being open.Water came out of the device.Unsure whether the device was being used for educational purposes.Not sure whether the water that came out was foreign matter or if it was part of the procedure.
 
Additional Manufacturer Narrative
H.6.Investigation summary: two samples model mfs141 were returned for investigation.The sets were examined for defects and abnormalities.No defects or abnormalities were observed.The sets were attached to a 10 ml bd syringe filled with water and attempted to flush while in the open position.The sets were successfully flushed with no leakage observed.The customer complaint that the devices would not allow fluid to flow through despite being open.Water came out of the device.Possible foreign matter leaking out could not be replicated.The root cause could not be determined because the issue could not be replicated.A device history record review for model mfs141 lot number 22039100 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 10mar2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXGUARD FLOW CONTROLLER EXTENSION SET WITH NEEDLELESS Y-SITE(S)
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15158709
Report Number9616066-2022-01039
Device Sequence Number12048553
Product Code FPA
UDI-Device Identifier10885403236471
UDI-Public10885403236471
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K051499
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/07/2025
Device Model NumberMFS141
Device Catalogue NumberMFS141
Device Lot Number22039100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2022
Supplement Date Received by Manufacturer09/07/2022
Initial Report FDA Received Date08/03/2022
Supplement Report FDA Received Date09/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-