|
MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
3545700 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
06/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a 36-year-old caucasian female patient underwent a breast augmentation revision surgery with 700cc mentor memorygel breast implants.Post-operatively, the patient was diagnosed with bilateral baker grade iii capsular contracture.As a result, the patient underwent explantation on (b)(6) 2022.Refer to manufacturing report number 1645337-2022-08983 for the contralateral event.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On august 30, 2022, the product was returned to mentor for evaluation.On september 06, 2022, mentor completed an evaluation on the returned device.Mentor conducted a visual inspection of the device.During visual evaluation, no apparent damage or visual anomalies were observed on the returned device.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsule, and effusion to pathology for examination.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|