WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G52918
Medical Device Problem Code Difficult or Delayed Activation (2577)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/27/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Description of event according to initial reporter: the physician placed filter, no deviation from how he always places them, from the right femoral vein approach.Message was that the red ¿lock¿ on the ivc filter wasn¿t working properly.The filter would not deploy and then it ended up deploying in the sheath.They tried to resheath it since it was only partially out of the sheath, however that just messed the secondary struts up.Ultimately, the physician got internal jugular access and had to snare the sheath to hold it in place while the tech slowly retracted the deployment sheath.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTER SET
Common Device Name
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key15159101
Report Number3005580113-2022-00087
Device Sequence Number12199485
Product Code DTK
UDI-Device Identifier10827002529189
UDI-Public(01)10827002529189(17)250218(10)E4225464
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/03/2022
Report Date (Section F) 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/18/2025
Device Model NumberG52918
Device Catalogue NumberIGTCFS-65-1-UNI-TULIP
Device Lot NumberE4225464
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date07/13/2022
Event Location Hospital
Date Report to Manufacturer08/03/2022
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/03/2022
Date Device Manufactured02/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Date Report Sent to FDA08/03/2022
-
-