Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 07/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient was revised approximately nine months post implantation due to implant fracture.There is no additional information available.
 
Additional Manufacturer Narrative
(b)(4).Report source- united kingdom.Customer has indicated that the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.No product was returned, or pictures provided; therefore, visual and dimensional evaluations could not be performed.Radiographs were provided and reviewed by a health care professional, and it was determined further review would not enhance the investigation.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Event or Problem Description
No additional information.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The reported event was able to be confirmed by examination of the returned product.Visual examination identified that the hinge post is fractured.The device was submitted for further analysis which determined that the returned device suspected to be fractured due to fatigue.The fracture surface showed smearing likely due to post fracture damage and biological debris, both of which resulted in obscuring the artifacts.Fatigue mode of fracture with step-like fracture artifacts observed in the nonsmeared areas.Material analysis show the implant is consistent with print specifications.Review of the device history records identified no deviations or anomalies during manufacruing.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE WITH SEGMENTAL HINGE POST SIZE B 12 MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15159459
Report Number0001822565-2022-02278
Device Sequence Number6167826
Product Code KRO
UDI-Device Identifier00889024195684
UDI-Public(01)00889024195684(17)290811(10)65154703
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K070978
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00585002012
Device Lot Number65154703
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2022
Supplement Date Received by Manufacturer10/13/2022
08/01/2023
Initial Report FDA Received Date08/03/2022
Supplement Report FDA Received Date10/25/2022
08/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age38 YR
Patient SexMale
Patient Weight140 KG
-
-