| Model Number |
8881513207 |
| Medical Device Problem Codes |
Break (1069); Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported that during inspection of lot 2209408764, several syringes were rejected for particulate matter on the plunger and broken syringes, leading to the lot failing inspection.A photo provided shows the breaks the customer is referring to are a broken luer tip, a cracked barrel, and a broken finger flange.
|
| |
|
Additional Manufacturer Narrative
|
|
There were four photos and twelve physical samples received for investigation.The samples were visually inspected, and the reported conditions were confirmed.One syringe had a broken tip and eleven syringes had particles in the plunger.The device history record was reviewed, and no abnormal process conditions were present during the manufacturing of the product that could have led to the reported conditions.The dhr review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.The product was manufactured on april 13, 2022.A gemba walk was performed at the manufacturing line and the current process and controls were found properly followed including finished product assembly, packaging, and inspections performed to the product.In addition, qc inspections are performed to the product for material release and production personnel performs a 100% visual inspection during the packaging process in order to detect and discard any identified issues.Based on all available information, a root cause related to the manufacturing process could not be determined.However, the appropriate production personnel did receive an alert notification to heighten awareness of this report.This complaint will be used for tracking and trending purposes.
|
| |
|
Search Alerts/Recalls
|