|
This follow-up is being submitted to relay additional information.Updated: b4, b5, g3, g6, h1, h2, h3, h6, and h10 visual examination of the returned device found signs of use (nicked and gouged) and the dovetail feature is flared and compressed.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.Damage to the articular surface can occur if the articular surface is not correctly placed and oriented before pushing it using the inserter instrument.Detailed instructions for inserting the articular surface are provided in persona the personalized knee surgical technique (97-5026-001-00, rev 13) pg.44-45.The root cause is attributed to use error.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|