| Model Number |
2760 |
| Medical Device Problem Code |
Failure to Zero (1683)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the pressure did not return to the initial pressure (0 atm).A encore 26 advantage kit was selected for use.During the procedure, the device was inflated at 26 atm.Upon deflation, the balloon was in negative pressure, however, the manometer remained at a pressure of 26 atm and did not return to the initial pressure of 0 atm.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: the encore device was returned for investigation.Analysis of returned product revealed that gauge needle was between 0 atm and 26 atm when received, consequently confirming the event reported.During functional testing, the yellow knob was difficult to activate.No further issues were noted.
|
| |
|
Event or Problem Description
|
|
It was reported that the pressure did not return to the initial pressure (0 atm).An encore 26 advantage kit was selected for use.During the procedure, the device was inflated at 26 atm.Upon deflation, the balloon was in negative pressure, however, the manometer remained at a pressure of 26 atm and did not return to the initial pressure of 0 atm.No patient complications were reported.
|
| |
|