Model Number IPN917286
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn#(b)(4).
 
Event or Problem Description
It was reported that: "cma educator and tm were running an iabp education session in icu on (b)(6) 2022 at (b)(6).During the session we noted that the condensate bottle at the back of the pump had a discoloured liquid in it.Notified staff from (b)(6)- asked to get a urine dipstick to check if it is blood.Upon dipstick test noted to be positive for blood.Staff unsure when the pump was last on a patient and nature of the incident".Additional information has been requested to the customer.No additional details have been provided.
 
Event or Problem Description
It was reported that:"cma educator and tm were running an iabp education session in icu on 13/5/22 at rmh.During the session we noted that the condensate bottle at the back of the pump had a discoloured liquid in it.Notified staff from rmh - asked to get a urine dipstick to check if it is blood.Upon dipstick test noted to be positive for blood.Staff unsure when the pump was last on a patient and nature of the incident".Additional information has been requested to the customer.No additional details have been provided.
 
Additional Manufacturer Narrative
(b)(4).The reported complaint of blood in the condensation bottle is not able to be confirmed.No iab rupture was reported to teleflex from the customer.The product was not returned for investigation.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC3 OPTIMUS IABP NA/AJLA
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section G)
ARROW INTERNATIONAL LLC
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
marling heaton
3015 carrington mill blvd
morrisville 27560
MDR Report Key15161095
Report Number3013162291-2022-00005
Device Sequence Number17604659
Product Code DSP
UDI-Device Identifier10801902172072
UDI-Public10801902172072
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberIPN917286
Device Catalogue NumberIAP-0701
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2022
Supplement Date Received by Manufacturer08/11/2022
Initial Report FDA Received Date08/03/2022
Supplement Report FDA Received Date08/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/28/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED; NONE REPORTED
Patient SexUnknown
-
-