| Model Number |
1804450-23 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat an ostial left main to the proximal left anterior descending artery.A 4.5x23mm xience skypoint was attempted to be delivered but failed to cross due to the stent implant getting caught on the proximal edge of a jailed balloon in the proximal left circumflex artery.Resistance met with the balloon during removal.The device was replaced with a non-abbott 4.0mm stent to continue the procedure.There was no adverse patient effects and no clinically significant delay.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other incidents.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling of the device.
|
| |
|