| Model Number |
HEMTOM10 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has not been returned for evaluation.When it arrives and the product evaluation has been completed, a supplemental report will be submitted.The device history record review was completed and all inspections passed with no non-conformances.
|
| |
|
Event or Problem Description
|
|
It was reported that the hemtom10 has been generating intermittent and fluctuating sto2 values.The issue has been with the a1 l values.The a2 r value stayed steady from 88 89 during a case.When there is a value the a1 l value fluctuated between 65 to 88 changing every few seconds.After a few minutes of displayed values the a1 l suddenly stops generating values.After several minutes the cycle starts again.This was during use.There is no patient injury.Patient demographics were requested but unavailable.A hem1 hemfsm10 and foresight sensor will also be sent in separate mdrs.
|
| |
|
Additional Manufacturer Narrative
|
|
One hemisphere tissue oximetry module was returned for evaluation.They connected the module to a known working system and began monitoring using sto2 simulators.They were able to use both the a and b ports with no issues.They ran each port for 30 minutes.The sto2 readings were accurate and remained steady.There was no defect found.
|
| |
|
Additional Manufacturer Narrative
|
|
Please refer to these submission numbers: 2015691-2022-07154 for the hemfsm10, 2015691-2022-07156 for the hemtom10, 2015691-2022-07152 for the hem1, and 2015691-2022-07153 for the foresight sensor.
|
| |
|