• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number MF-410640
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 12/03/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: "concern with the product," "a lump on left breast", "left transaxiliary is numb", "chest pain", "feels something moving down left arm", capsular contracture baker grade unknown.
 
Event or Problem Description
Patient reported exchange of textured breast implants due to concern with the product, "a lump on left breast", "left transaxiliary is numb", "chest pain", and "feels something moving down left arm." patient additionally reported "cancer survivor"; this is not device related.Healthcare professional later reported capsular contracture baker grade unknown.Device has been explanted.
 
Event or Problem Description
Patient reported exchange of textured breast implants due to concern with the product, "a lump on left breast", "left transaxiliary is numb", "chest pain", and "feels something moving down left arm." patient additionally reported "cancer survivor"; this is not device related.Healthcare professional later reported capsular contracture baker grade unknown.Device has been explanted.
 
Additional Manufacturer Narrative
Correction: aware date of initial medwatch should have been 11/jul/2022.This was date we first became aware of reportable event and matches the parent record aware date.Device evaluation: the device related to the reported event of anxiety-product/procedure, lump/nodule, pain, sensation increase/decrease, neurological symptoms and capsular contracture was received on july 29, 2022, with lot number 2830357.Analysis of the returned device identified the weight the device within specification, red particles in the shell, crease fold and encapsulation.Cloudy and voids in the gel observed after autoclave cycle.Based on the device analysis the final assessment is: no issues found related with the manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15161344
Report Number9617229-2022-13694
Device Sequence Number5812305
Product Code FTR
UDI-Device Identifier10888628003521
UDI-Public10888628003521
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/12/2020
Device Model NumberMF-410640
Device Catalogue NumberMF-410640
Device Lot Number2830357
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/13/2019
Supplement Date Received by Manufacturer08/09/2022
Initial Report FDA Received Date08/03/2022
Supplement Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/12/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexFemale
-
-