| Model Number |
41488 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the coating unraveled.Two 0.035in, 185cm, j tip back-up meier guidewires were used in a tavr procedure.While advancing the guidewire through the sheath, the guidewire's coating unraveled.This occurred on both devices.A non-boston scientific guidewire was used to complete the procedure.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Media analysis by manufacturer: the customer provided a picture.Per media inspection, it can be observed that the ptfe of the guidewire was peeled off.
|
| |
|
Additional Manufacturer Narrative
|
|
Media analysis by manufacturer: the customer provided a picture.Per media inspection, it can be observed that the ptfe of the guidewire was peeled off.Device eval by manufacturer: the returned guidewire was bent/kinked at the distal end.Additionally, visual and microscopic inspection revealed that the coating ptfe was peeled off at several sections.
|
| |
|
Event or Problem Description
|
|
It was reported that the coating unraveled.Two 0.035in, 185cm, j tip back-up meier guidewires were used in a tavr procedure.While advancing the guidewire through the sheath, the guidewire's coating unraveled.This occurred on both devices.A non-boston scientific guidewire was used to complete the procedure.No patient complications were reported.
|
| |
|