• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HLS SET ADVANCED; OXYGENATOR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HLS SET ADVANCED; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BE-HLS 7050 #SHLS SET ADVANCED 7.0
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/16/2022
Event Type  malfunction  
Manufacturer Narrative
It was reported that bloody condensation water emerged from the gas outlet of the oxygenator.No harm to any person has been reported.The affected hls set was not available for further investigation, as it was discarded by the customer.Therefore no probable root cause could be determined.A review of the trend data for similar complaints was performed and no similar complaints were found with a root cause.Based on the provided video the reported failure "leakage gas outlet" could be confirmed.The production records of the affected hls module were reviewed on 2022-08-02.Following tests are performed according to the bop as a 100 % inspection: pressure test heat exchanger; leak test water/gas side; pressure test blood side; final functional test.According to the final test results, all hls modules passed the tests as per specifications.Production related influences are unlikely to have contributed to the reported failure.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
It was reported that bloody condensation water emerged from the gas outlet of the oxygenator.No harm to any person has been reported.Complaint id #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLS SET ADVANCED
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key15162486
MDR Text Key304924098
Report Number8010762-2022-00300
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeCS
PMA/PMN Number
K112360
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBE-HLS 7050 #SHLS SET ADVANCED 7.0
Device Catalogue Number701069073
Device Lot Number3000207497
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/18/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age24 YR
Patient SexPrefer Not To Disclose
-
-