| Model Number |
ILRGHG |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the abutment screw fractured at the site.Abutment screw was removed and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number: (b)(4).Additional 510(k) number is k072642.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).The reported device was returned for investigation.Visual evaluation of the returned abutments identified fractures at threads.No pre-existing conditions were noted on the product experience report (per).Patient was reported to have moderate density - (type ii) bone and the reported device was located on and tooth sites #25.Complainant reported that the abutment screw fractured at the site.The reported complaint was confirmed.A definitive root cause for this complaint could not be determined.A device history record (dhr)review was performed and no related deviations or nonconformance's were noted.Also, a complaint history search was performed using our complaint handling system and there were no related complaints for this product lot.The following sections have been updated: h3: device evaluated by manufacturer; h6: evaluation codes; h10: additional narrative.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Search Alerts/Recalls
|