(b)(4).Concomitant medical products: 0100551207, fitmoreâ®, hip stem, uncemented, b/7, taper 12/14, 3071654.00877503602, bioloxâ® delta, ceramic femoral head, m, 㸠36/0, taper 12/14, 3090759.Customer has indicated that the product will not be returned to zimmer biomet for investigation, requested but not returned by hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
This follow-up report is being submitted to relay additional information.Radiographs were provided and reviewed by a health care professional.Visual examination of the provided pictures identified the head, liner, and stem were explanted and are covered in bio-debris.No further evaluation can be made from the provided picture.Review of the device history records identified no deviations or anomalies during manufacturing.It was reported the patient dislocated after a fall.However, as the reason for the fall is unknown, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|