Model Number 0220210000
Medical Device Problem Code Loss of Power (1475)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was loss of light.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Additional Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.The technical service report is attached (see comm log), and indicates: examination description: at the time of the visit, there was no evidence of a problem with the 1288 arthroscopy system's monitor screen flickering, but the charge nurse and me reported that the recording device shut down (turned off) during the procedure.Work performed: checked the disconnection of the wiring of the video route, checked the power connection and operation of each device, and checked the operation of the recording device (sony's hvo-3300mt).Results : no problems were found with the 1288 arthroscope system screen, but just to be safe, the amperage of each device plugged into the outlet on the endcart was checked.Checked the amperage value of each device plugged into the end-cart outlet, and plugged in each device again so that the capacity does not exceed, and then plugged in the over capacity separately.We then contacted sony to arrange for a replacement recorder.Probable root cause: light cable, optics, leds, main board, jaw assembly, bent esst contact, inconsistent esst contact, cable pull out, camera head, firewire cable, software, front board, power supply, thermal switch, ac inlet board, ac inlet filter, touch screen, workmanship, inadequate air flow , inadequate calibration, excessive lint buildup inside the unit, use errors.The reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that there was loss of light.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L9000 LIGHTSOURCE INTERNATIONAL KIT
Common Device Name
LIGHT SOURCE, FIBEROPTIC, ROUTINE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15163115
Report Number0002936485-2022-00446
Device Sequence Number15076780
Product Code FCW
UDI-Device Identifier07613327051339
UDI-Public07613327051339
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K082813
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0220210000
Device Catalogue Number0220210000I
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/29/2022
Supplement Date Received by Manufacturer07/29/2022
Initial Report FDA Received Date08/04/2022
Supplement Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-