| Catalog Number |
3910-500-391 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the eyelet broke during the procedure.All pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: implant broken.Probable root cause: design: inadequate raw material selection for eyelet anchor; design geometry too sensitive to withstand clinical impaction loads; eyelet/inserter interface not designed to maintain proper assembly.Process: eyelet anchor or inserter shaft not manufactured to specification; incorrect material used during manufacturing; eyelet anchor and inserter not assembled to specification.Application: excessive force or leveraging of the anchor against the bone; user fully seats eyelet anchor onto inserter which eliminates interference fit; inadequate assessment of bone quality, pilot hole not prepared.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
|
| |
|
Event or Problem Description
|
|
It was reported that the eyelet broke during the procedure.All pieces were retrieved.
|
| |
|