Model Number 21-2961-24
Medical Device Problem Codes Use of Device Problem (1670); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Post Operative Wound Infection (2446)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during the use of the product, the skin at the puncture site became suppurated.Additionally the reporter stated "i recognize that it is a problem with the technique, but i would like you to analyze the product as well for good liquid flow.".
 
Additional Manufacturer Narrative
H10: no lot number was provided; therefore, device history record review could not be performed.A product sample was received for evaluation.Visual and functional testing was performed.Visual inspection found the sample was received in used condition and activated, and no issues was observed.Functional testing could not be performed since the sample was received in used condition and activated.Root cause analysis can be performed since the failure reported by the customer is not clear and failure could not duplicate in the manufacturing process since the product pass only for two kind of inspections ? visual inspection ? and ?occlusion test ?.This issue will be monitored for any increase in occurrence.If the occurrence of this issue increases significantly, additional investigative action will be taken accordingly.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Additional Manufacturer Narrative
Other, other text: this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).Please disregard the initial report submitted with the mfr number: 3012307300-2022-14629.The report was submitted in error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTEC GRIPPER PLUS NEEDLE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
3-10-15, hakata station east,
minneapolis, MN 55442
MDR Report Key15163218
Report Number3012307300-2022-14629
Device Sequence Number14929086
Product Code FPA
UDI-Device Identifier10610586027697
UDI-Public10610586027697
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K021999
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number21-2961-24
Device Catalogue Number21-2961-24
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/05/2022
Supplement Date Received by Manufacturer12/22/2022
06/20/2023
Initial Report FDA Received Date08/04/2022
Supplement Report FDA Received Date04/27/2023
06/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-