| Model Number |
21-2961-24 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
Post Operative Wound Infection (2446)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during the use of the product, the skin at the puncture site became suppurated.Additionally the reporter stated "i recognize that it is a problem with the technique, but i would like you to analyze the product as well for good liquid flow.".
|
| |
|
Additional Manufacturer Narrative
|
|
H10: no lot number was provided; therefore, device history record review could not be performed.A product sample was received for evaluation.Visual and functional testing was performed.Visual inspection found the sample was received in used condition and activated, and no issues was observed.Functional testing could not be performed since the sample was received in used condition and activated.Root cause analysis can be performed since the failure reported by the customer is not clear and failure could not duplicate in the manufacturing process since the product pass only for two kind of inspections ? visual inspection ? and ?occlusion test ?.This issue will be monitored for any increase in occurrence.If the occurrence of this issue increases significantly, additional investigative action will be taken accordingly.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).Please disregard the initial report submitted with the mfr number: 3012307300-2022-14629.The report was submitted in error.
|
| |
|
Search Alerts/Recalls
|