• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION VISTA 500

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION VISTA 500 Back to Search Results
Model Number DIMENSION VISTA 500
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
Two discordant, falsely depressed carbon dioxide patient samples (2 patients) were obtained on a dimension vista 500 instrument.The initial results were reported to the physician(s).The same patient samples were reprocessed on the same instrument.The repeat results were reported, as the correct results, to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the falsely depressed carbon dioxide patient samples.
 
Additional Manufacturer Narrative
A united states (us) customer reported two discordant, falsely depressed carbon dioxide patient samples (2 patients) obtained on a dimension vista 500 instrument.The customer contacted the siemens customer care center (ccc) and a siemens customer service engineer (cse) was dispatched to the customer's site and inspected the instrument.The cse replaced the sample 1 mixer which resolved the issue.Quality control was within acceptable limits after service and the instrument was fully operational upon departure.The instrument is performing according to specifications.No further evaluation of this device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIMENSION VISTA 500
Common Device Name
DIMENSION VISTA 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
500 gbc drive
po box 6101
newark DE 19714 6101
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
101 silvermine road
registration # : 1226181
brookfield CT 06804
Manufacturer Contact
eric so
511 benedict ave
tarrytown, NY 10591
9145243067
MDR Report Key15163303
Report Number2517506-2022-00194
Device Sequence Number14920550
Product Code JJE
UDI-Device Identifier00630414593517
UDI-Public00630414593517
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K051087
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDIMENSION VISTA 500
Device Catalogue Number10488217
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/11/2022
Initial Report FDA Received Date08/04/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/11/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-