• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD DEHP FREE 4 IN EXT SET W/ANTSPHN VLV DET; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD DEHP FREE 4 IN EXT SET W/ANTSPHN VLV DET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number C25012
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bd dehp free 4 in ext set w/antsphn vlv det separated during use.The following information was provided by the initial reporter: the anti syphon line came apart at the middle connection and the chemotherapy line hit the ground.
 
Additional Manufacturer Narrative
Investigation summary: no product or photo was returned by the customer.The customer complaint of the anti-syphon line coming apart at the middle connection could not be verified due to the product not being returned for failure investigation.Device history record review for model c25012 lot number 21059533 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 23jun2021.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.The root cause of this failure could not be identified without a failure investigation.The reported set is a simple extension set, and has no hand tightened connections that could have been an issue.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD DEHP FREE 4 IN EXT SET W/ANTSPHN VLV DET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
ryan reynolds
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15163372
Report Number9616066-2022-01044
Device Sequence Number12385661
Product Code FPA
UDI-Device Identifier50885403235618
UDI-Public50885403235618
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K790108
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/24/2024
Device Model NumberC25012
Device Catalogue NumberC25012
Device Lot Number21059533
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/05/2022
Initial Report FDA Received Date08/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-