Stent moved into bladder."as per cc form": insertion of the stent was problem - free on (b)(6) 2022, on (b)(6) the patient reported kidney pain.The ultrasound showed that the stent is not place correctly.In addition the kidney was third - degree congested.Lab complete (b)(6) 2022: visual inspection stent returned.Slight encrustation observed on one of the pigtail curls.Pigtail curl on the other end of the stent slightly raised.No other defects observed.Functional inspection n/a.A section of the device did not remain inside the patient¿s body.The patient did require any additional procedures due to this occurrence.New stent insertion.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.1.Are any images available for review? n/a, yes, no.2.What was the target location for the stent? one pigtail in the kidney and the other in the bladder.3.Did anything have to be removed from the patient? n/a, yes, no.4.If yes, please specify: resonance stent.(i) what part of the body the device was removed from? bladder.(ii) what device was removed? resonance stent.(iii) what instrument was used to remove it? cystoscope.5.Please specify the storage conditions of the device at the facility, particularly those relating to light and temperature use or return product, no storage at the hospital.6.Why was the stent removed? n/a, exchange to an other stent but not resonance.Patient kidney had a third-degree congestion.7.If other, please detail why the stent was removed.8.If the stent was removed what was used to complete the procedure? silicone stent.9.What was the length of the indwell time? 3 weeks.10.What disease mode was the physician trying to treat? na.11.What type and size wire guide was used with the device? 0.038 inch.12.Did the device come with a pigtail straightener? n/a, yes, no.13.If yes, was the pigtail straightener used? n/a, yes, no.14.Did the user attempt to attach the stent to the inserter before or after wire guide insertion? n/a, before, after the stent need no wire for insertion, it has no open end.15.Did the user attempt to attach the tapered end of the stent to the inserter? n/a, yes, no, both ends are the same.16.Was this device assembled outside of the body?n/a, yes, no.17.Did the patient require any additional procedures as a result of this event? n/a, yes, no.18.What intervention (if any) was required? new stent insertion.19.Was the secondary intervention performed during the same procedure as the device failure or was it scheduled for another day? n/a, same procedure, another day.20.Were any other defects (other than the complaint issue) observed on the device prior to return (e.G.Kink)? n/a, yes, no.21.If yes, please specify what was observed and where on the device it was observed.Rep update 25/07/2022: 1.Was there difficulty advancing the stent to the target location? n/a, yes, no.2.Please describe the purpose of stent placement/ disease mode the surgeon was trying to treat.Extrinsic ureteral obstruction.3.Are any images available for review? n/a, yes, no.4.What is the patient¿s anatomy (calcified, tortuous)? n/a, calcified, tortuous, both.5.What is the patient¿s status or outcome? patient is fine, he got a silicone stent.6.Describe the method used to confirm stent placement normally resonance stent placing.7.Was the patient using calcium supplementation? n/a, yes, no.8.Was force required to remove the stent? n/a, yes, no, the stent was in the bladder.9.Was encrustation evident on the stent? n/a, yes, no, looked totally normal.
|
|
Device evaluation: the rms-060026-r device of lot number c1924491 involved in this complaint was returned for evaluation, without the original packaging.With the information provided, a physical examination and document-based investigation was conducted.Lab evaluation: the device related to this occurrence underwent a laboratory evaluation on the 20th july 2022.On evaluation of the device the stent was observed with encrustation on one of the pigtail curls.The pigtail on the other end of the stent was slightly raised.No other defects were observed.Please note that the device was returned without the original packaging in a plastic bag without padding.It is therefore probable that the end of the pigtail on the stent may have become slightly bent due to impact on transportation.Document review: prior to distribution rms-060026-r devices are subjected to a visual inspection and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl a review of the manufacturing records for rms-060026-r of lot number c1924491 did not reveal any discrepancies that could have contributed to this complaint issue.It should be noted, there is two nc¿s noted on the work order however, there is no sufficient evidence to suggest this impacts the complaint device, as the affected products were re-worked.The review of relevant manufacturing records confirms the failure mode has not previously occurred for this work order.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1924491.It should be noted that the instructions for use (ifu0020) note migration or stent dislodgement as a potential adverse event associated with the use of the rms device.The ifu also states as follows: ¿individual variations of interaction between stents and the urinary system are unpredictable.Use of this device should be based upon consideration of risk-benefit factors as they apply to your patient.Informed consent should be obtained to maximize patient compliance with follow-up procedures.¿ there is no evidence to suggest that the customer did not follow the instructions for use.Root cause review a definitive root cause could not be determined.A possible root cause could be attributed to inherent risk of the device, as migration or stent dislodgement is a potential adverse event associated with the use of the rms device.Another possible root cause could be attributed to possible difficult patient anatomy.Difficult anatomy may have compressed the indwelling device, placing excessive force on the stent.This may have caused or contributed to device migration.Summary: migration is the failure mode identified for this event.A possible root cause of inherent risk of device and difficult patient anatomy was identified.Complaint is confirmed based on customer and/or rep testimony.According to the initial reporter, the patient required surgical intervention to retrieve the migrated stent due to the occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
|
|
Device evaluation: the rms-060026-r device of lot number c1924491 involved in this complaint was returned for evaluation, without the original packaging.With the information provided, a physical examination and document-based investigation was conducted.Lab evaluation: the device related to this occurrence underwent a laboratory evaluation.On evaluation of the device the stent was observed with encrustation on one of the pigtail curls.The pigtail on the other end of the stent was slightly raised.No other defects were observed.Please note that the device was returned without the original packaging in a plastic bag without padding.It is therefore probable that the end of the pigtail on the stent may have become slightly bent due to impact on transportation.Document review: prior to distribution rms-060026-r devices are subjected to a visual inspection and functional checks to ensure device integrity.A review of the manufacturing records for rms-060026-r of lot number c1924491 did not reveal any discrepancies that could have contributed to this complaint issue.It should be noted, there is two nc¿s noted on the work order however, there is no sufficient evidence to suggest this impacts the complaint device, as the affected products were re-worked.The review of relevant manufacturing records confirms the failure mode has not previously occurred for this work order.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1924491.It should be noted that the instructions for use (ifu0020) note migration or stent dislodgement as a potential adverse event associated with the use of the rms device.The ifu also states as follows: ¿individual variations of interaction between stents and the urinary system are unpredictable.Use of this device should be based upon consideration of risk-benefit factors as they apply to your patient.Informed consent should be obtained to maximize patient compliance with follow-up procedures.¿ there is no evidence to suggest that the customer did not follow the instructions for use.Image review: an image was not returned for evaluation.Root cause review: a definitive root cause could not be determined.A possible root cause could be attributed to inherent risk of the device, as migration or stent dislodgement is a potential adverse event associated with the use of the rms device.Another possible root cause could be attributed to possible difficult patient anatomy.Difficult anatomy may have compressed the indwelling device, placing excessive force on the stent.This may have caused or contributed to device migration.Summary: migration is the failure mode identified for this event.A possible root cause of inherent risk of device and difficult patient anatomy was identified.Complaint is confirmed based on customer and/or rep testimony.According to the initial reporter, the patient required surgical intervention to retrieve the migrated stent due to the occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
|