• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ULTRATHANE AMPLATZ UNIVERSAL DRAINAGE CATHETER; GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ULTRATHANE AMPLATZ UNIVERSAL DRAINAGE CATHETER; GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number N/A
Medical Device Problem Codes Deformation Due to Compressive Stress (2889); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a patient required replacement of an ultrathane amplatz universal drainage catheter.On (b)(6) 2022, the drainage catheter was initially placed without difficulty.The patient was described as active postoperatively.One week after placement, on (b)(6) 2022, a kink at the proximal side of the drainage catheter was found.The catheter was replaced with a new catheter successfully.No other adverse effects were reported for this incident.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.Additional customer information: phone: (b)(6).Pma/510(k) #: exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation on 25jul2022, a representative of (b)(6) hospital (japan) reported to cook that an ultrathane amplatz universal drainage catheter (rpn: ult12.0-38-50-p-6s-audc-pig; lot#: 9968860) kinked.The device was required for a percutaneous transhepatic cholangiography (ptcd) procedure on (b)(6) 2022 and was placed in the patient without difficulty.About a week later, on (b)(6) 2022, a kink at the proximal side of the drainage catheter was observed.As a result, surgical intervention was required to replace the catheter to prevent permanent harm.No other adverse effects were reported.Reviews of documentation including the complaint history, device history record (dhr), quality control procedures, manufacturing instructions (mi), and instructions for use (ifu) of the complaint device were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot 9968860 and the related subassembly lots revealed no related non-conformances.To date, a further search of our database records found this to be the only complaint received involving the reported lot number.Since there is objective evidence the dhr was fully executed, cook medical inc.Has concluded there is no evidence the devices were manufactured out of specification, or that there are non-conforming devices in house or in the field.Cook also reviewed product labeling.The product ifu, t_multi_rev4, provides the following information to the user related to the reported failure mode: precautions: "a tfe-coated wire guide must be used with ultrathane® catheters " how supplied "upon removal from package, inspect the product to ensure no damage has occurred." based on the information provided, no product returned, and the results of the investigation, cook medical has concluded the cause to be component failure without any design or manufacturing issue.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE AMPLATZ UNIVERSAL DRAINAGE CATHETER
Common Device Name
GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key15164121
Report Number1820334-2022-01309
Device Sequence Number17575821
Product Code GBO
UDI-Device Identifier00827002119758
UDI-Public(01)00827002119758(17)220823(10)9968860
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/23/2022
Device Model NumberN/A
Device Catalogue NumberULT12.0-38-50-P-6S-AUDC-PIG
Device Lot Number9968860
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/25/2022
Supplement Date Received by Manufacturer03/07/2023
Initial Report FDA Received Date08/04/2022
Supplement Report FDA Received Date03/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/23/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-