• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, FRENCH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, FRENCH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3013-64
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event or Problem Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) had a display failure.There was no harm to the patient.
 
Additional Manufacturer Narrative
During a preventive maintenance (pm) at customer's site, the getinge field service engineer (fse) was not able to reproduce the reported issue.The fse reconnected the video cable / contact on the cable.The machine is functional.The fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.Attempts were made to find out if the iabp was released to the customer and cleared for clinical service but no response.If information is received, we will reopen and update the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 INTRA-AORTIC BALLOON PUMP, FRENCH, 220V
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15164511
Report Number2249723-2022-01936
Device Sequence Number2123454
Product Code DSP
UDI-Device Identifier10607567107509
UDI-Public10607567107509
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K063525
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0998-00-3013-64
Device Catalogue Number0998-00-3013-64
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2022
Supplement Date Received by Manufacturer03/27/2023
05/24/2023
Initial Report FDA Received Date08/04/2022
Supplement Report FDA Received Date05/01/2023
05/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/31/2001
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-