MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number DBE-150
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis of the reported device is in progress.A supplemental report will be submitted when the device analysis is completed.Csi id: (b)(4).
 
Event or Problem Description
While preparing the stealth 360 peripheral orbital atherectomy device (oad) to reenter the patient for a second treatment, the user noted a pinhole leak located near the mid shaft of the saline sheath.The flow of the saline to the tip of the oad was considered sufficient to avoid injury to the patient, however, a new oad was used to complete the procedure.The patient was stable.
 
Additional Manufacturer Narrative
The device was returned to csi.Analysis revealed damage to the saline sheath 33.5cm from the distal end of the sheath.During functional testing, fluid was observed leaking from the damaged site.The root cause of the saline sheath damage is inconclusive.The device history record for the reported oad could not be reviewed as the lot number was not provided.The results of the investigation are inconclusive since the reported device was not returned for analysis.Csi id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
tonia moskalets
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key15165008
Report Number3004742232-2022-00186
Device Sequence Number4622984
Product Code MCW
UDI-Device Identifier10850000491332
UDI-Public(01)10850000491332(17)231130(10)411894-1
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model NumberDBE-150
Device Catalogue Number7-10059-08
Device Lot Number411894-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2022
Supplement Date Received by Manufacturer07/21/2022
Initial Report FDA Received Date08/04/2022
Supplement Report FDA Received Date08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexMale
Patient Weight81 KG
Patient RaceWhite
-
-