| Model Number |
10464344 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported that they noticed a shift low in results when doing patient correlations testing for cancer antigen 15-3 (ca15-3) on the dimension vista® 500 system, serial number (b)(4).Patient correlation sample results were not reported to physicians.There are no known reports of patient intervention or adverse health consequences due to the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens healthcare diagnostics inc.Has investigated the event of a negative bias with the dimension vista loci ca 15-3 assay on the dimension vista system.Siemens has confirmed the dimension vista loci ca 15-3 assay showed a negative bias with patient samples.Siemens issued an urgent medical device correction, vc-22-02.A.Us dated (b)(6) 2022 and an urgent field safety notice vc-22-02.A.Ous dated (b)(6) 2022 to all dimension vista loci ca 15-3 customers.The letters instructed customers to use lot specific correlation factors provided for the affected lots.Siemens healthineers is working to restore assay performance and until such time, subsequent lots of dimension vista loci ca 15-3 may contain an alert card within the carton listing lot specific correlation factors.
|
| |
|
Search Alerts/Recalls
|