• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH SINGLE-INNER SETSCREW; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH SINGLE-INNER SETSCREW; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 179702000
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/10/2022
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Procode: additional device product codes: kwp;kwq;mnh;mni;osh.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a j&j employee.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from depuy synthes reports an event in colombia as follows: it was reported that during a procedure on (b)(6) 2022, a screw was inserted in the pedicle, bar and top were placed without torsion.During tightening the internal screw did not lower and it was twisted, so another internal screw was tried and the same happened.The bar was removed and it was observed that the screw had the head with the internal part twisted which prevented the correct closing of the system.In the tightening operation with the new screw, the screwdrivers were also twisted.The surgery was completed without further inconvenience and without affecting the patient.This report is for one (1) single-inner set screw.This is report 4 of 6 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6: the subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: a product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the single-inner setscrew was found the device stripped on thread en inner part of the screw.No other defect was found.A dimensional inspection for the single-inner setscrew was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the single-inner setscrew would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h4, h6: a manufacturing record evaluation was performed for the finished device: product code: 179702000.Lot number: 337534.It was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 07.03.2022.Qty: (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE-INNER SETSCREW
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle 02400
SZ   02400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
8472871282
MDR Report Key15165499
MDR Text Key304920153
Report Number1526439-2022-01348
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034064307
UDI-Public(01)10705034064307
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K160904
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number179702000
Device Catalogue Number179702000
Device Lot Number337534
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received09/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2022
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
SI POLYAXL SCREW 7 X 50MM; SINGLE-INNER SETSCREW; SINGLE-INNER SETSCREW; X25 FINAL TIGHTENER; X25 FINAL TIGHTENER
-
-