GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) AESPIRE 7100; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Increase in Pressure (1491)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/26/2022
Event Type  malfunction  
Event Description
Per china aer database: it was reported that during patient use the anesthesia unit delivered high airway pressure and the manual ventilation is described as "not good".The patient was transferred to a different device and there was no report of patient harm.
 
Manufacturer Narrative
The unit was evaluated by on-site biomedical engineer and found the flow sensor is disabled and damaged.Further evaluation found the flapper of the flow sensor was stuck to the top of the flow sensor housing.The flow sensor was replaced to resolve the issue.Patient information could not be obtained due to country privacy laws.The initial reporter is located outside the u.S.And therefore this information is not provided due to country privacy laws.Legal manufacturer: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESPIRE 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH  214028
Manufacturer (Section G)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH   214028
Manufacturer Contact
anthony amenson
MDR Report Key15168741
MDR Text Key302785833
Report Number9710602-2022-00426
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial
Report Date 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-