| Model Number |
115-150 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
06/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient reported rupture against left side device.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.
|
| |
|
Event or Problem Description
|
|
Patient reported rupture against left side device.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed a broken sharp assessed as unidentified (tear) opening (shell thickness was within specification).Additional observations: crease flat observed in the device.Wear abrasion observed in the device.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Patient reported rupture against left side device.The device has been explanted and replaced.
|
| |
|