| Model Number |
10-0000-07 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The service visit at kaiser permanente regional reference lab - (b)(4) was scheduled for a 12 month pm within work order: (b)(4).During the site visit, the hybridization (hyb) heater required recalibration due to failing initial verification at setpoint 2.Field quality updated risk assessment questions 7 and 8 were updated to 'yes' since the setpoint temperature failed outside the accepted limits established by the verigene i assay temperature tolerance guard band risk report (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
In review of all associated customer-provided documentation and internally generated investigation notes against the criteria set forth in document (b)(4) (medical device reporting standards operating procedure), this complaint does meet the requirements for an mdr in the countries where this product is sold or distributed.Filing individual mdr out of abundance of caution due to the remote possibility of a false result contributing to a death or serious injury.
|
| |
|