| Model Number |
10-0000-07R |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The service visit at blanchard valley health system was scheduled after the customer reported no calls using the bc-gp and bc-gn assay within work order wo-00199347.During the site visit, the amplification (amp) heater required recalibration due to failing initial verification at setpoint 3.Field quality updated risk assessment questions 7 and 8 to 'yes' since the setpoint temperatures failed outside the accepted limits established by the verigene i assay temperature tolerance guard band risk report (03838).
|
| |
|
Additional Manufacturer Narrative
|
|
In review of all associated customer-provided documentation and internally generated investigation notes against the criteria set forth in document 00043 (medical device reporting standards operating procedure), this complaint does meet the requirements for an mdr in the countries where this product is sold or distributed.Filing individual mdr out of abundance of caution due to the remote possibility of a false result contributing to a death or serious injury.
|
| |
|
Search Alerts/Recalls
|