| Model Number |
234-102-065 |
| Medical Device Problem Code |
Device Misassembled During Manufacturing /Shipping (2912)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product was returned for evalution.A review of the device history record was performed and all product met requirements prior to release.A visual of the affected device was sent to riverpoint for analysis.The analysis concluded an error may have occured during manufacturing assembly of the device.The lot retains underwent visual inspection and all product passed inspection.This report and use of categorical definitions required by fda 3500a does not constitute an admission by riverpoint medical or its employees that riverpoint medical or its employees have caused or contributed to the event described in this report.Riverpoint medical has filed this information to comply with the medical device reporting regulation 21 cfr 803.If additional information is provided to riverpoint medical regarding this event, a supplementary 3500a form will be submitted as required by the fda.
|
| |
|
Event or Problem Description
|
|
According to the reporter, "the pa constructed the open loop procinch at the back table following all manufacturers instructions, and when they pulled the graft down and away from the card, the cinching mechanism wouldn't pull into the center holes of the procinch button.One of the loops was through the button, and the other did not pass through any holes on the button.Both white sutures creating loops should be through center holes of the button.".
|
| |
|