| Model Number |
EQUINOXE REVERSE 38MM HUMERAL LINER +2.5 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
07/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
This 76 year old, male patient presented to the hospital with a dislocated right shoulder.The patient consented to revision surgery, having previously had a total shoulder replacement for trauma.The surgeon discovered that the trauma stem was retroverted more than usual, which explained the dislocation.Microbiology samples were taken and the 0mm humeral tray and 0mm humeral liner were removed and replaced with a new 0mm humeral tray and a 5mm humeral liner.The surgeon was unable to reattach the greater trochanter, so left it.The shoulder was stable and secure.There was no reported breakage of devices or surgical delay/prolongation.The patient was last known to be instable condition following the event.No other patient information/medical history reported.The devices are not returning as the hospital will not allow implants to be desterilized for return.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending evaluation.320-10-00, 7318926 - equinoxe reverse tray adapter plate tray +0.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3)based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of the surgical event cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: g2, h6, h11.No device was returned for evaluation; the reported event was confirmed by review of radiograph images provided.The review identified that the joint appears to be dislocated.Additionally, the proximal portion appears to be fractured.This is consistent with the reported information of a prior trauma and likely contributed to the dislocation.Operative notes and/or medical records were not provided for review of usage/ technique.A review of complaint history determined that no further action is required as no trends were identified.A definitive root cause was unable to be determined; however, may have resulted from the patient's fractured humerus from a previous trauma.Dislocation is a known risk, as outlined in the ifu.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The following section has been added: h6 - f code.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|